Introduction:
Standard parameters: ECG SPO2 NIBP RESP TEMP PR
Optional parameters: 2-IBP CO2 thermal printer.
Features:
Display mode: 12'' color TFT LCD with high-resolution.
Be applicable for adult, pediatric and neonatal for all-round monitoring.
Operation interface with Chinese and English. (Optional languages: French, German, Italian, Turkish, etc.)
Multiple display modes are optional, such as big characters interface.
Waveform and data colour can be set optionally.
Storage of 480-hour trend data, and review of 40-second holographic waveform.
7-lead ECG displays in a screen, and ECG waveforms series display.
Anti-high frequency surgical unit, defibrillation-proof(requirement for special leads).
Adopt digital SpO2 technology, which has strong anti-interference and anti-weak filling capability.
Function of NIBP review, storage for up to 400 NIBP data.
Connect central monitoring station, other bed observation and software updating with wireless and wired mode.
Built-in SD card, U disk for storing data and U disk updating.
Support the functions of VGA video output and adjustable LCD backlight.
Support alarm with human voice body.
Function of synchronous signal output.
Three operation modes: monitoring, surgery and diagnosis.
Auto-analysis for arrhythmia waveform.
Calculation function for drug concentration and titration table.
Dual-channel temperature monitoring, and with temperature error display.
NIBP dual-overpressure protection with software and hardware.
Built-in thermal printer.
Built-in rechargeable lithium battery for uninterrupted monitoring.
The product has passed FDA certification in America and CE certification in Europe.
Performance:
ECG
Lead Mode 3-lead and 5-lead are optional
Lead Selection I, II, III, avR, avL, avF, V
Gain Selection ×0.25, ×0.5, ×1, ×2
HR Range 15 ~ 300 bpm
Measuring and Alarm Range for ST Segment -0.6 mV~ + 0.8 mV
Scan Speed for ECG Waveform 12.5mm/s, 25mm/s, 50mm/s
RESP
Measuring Method R-F(RA-LL) Impedance
Measuring and Alarm Range for Resp. Rate:
Adult 0~120rpm
Pediatric and Neonatal 0~150rpm
Apnea Alarm 10 ~ 40 S
SpO2
Measuring Range 0 ~ 100 %
Accuracy
70% ~ 100% ±2%
0% ~ 69% unspecified
Measuring range for pulse rate 0~250bpm
SpO2 probe SpO2 probe for adult, pediatric and neonatal are optional
NIBP
Method Oscillometry
Mode Manual, Auto, Continuous
Measuring Period in Continuous Mode 5 Min
Measuring and Alarm Range 10~270mmHg
Overpressure protection dual-protection with software and hardware
Cuff cuff for adult, pediatric and neonatal are optional
TEMP
Channel dual-channel
Measuring and Alarm Range 0 ~ 50°C
Resolution 0.1°C
Accuracy ±0.1°C
EtCO2(optional)
Method Sidestream or Mainstream
Measuring Range for CO2 0~150mmHg
Resolution for CO2:
0.1 mm Hg 0 to 69 mm Hg
0.25 mm Hg 70 to 150 mm Hg
Accuracy for CO2:
0 - 40 mm Hg ±2 mm Hg
41 - 70 mm Hg ±5%
71 - 100 mm Hg ±8%
101 - 150 mm Hg ±10%
Respiration Rate>80BPM ±12%
AwRR Range 2~150 rpm
AwRR Accuracy ±1BPM
Apnea Alarm Available
IBP(optional)
Channel dual-channel
Label ART, PA, CVP, RAP, LAP, ICP, P1, P2
Measuring and Alarm Range -50~350 mm Hg
Resolution 1 mm Hg
Accuracy ±2% or 1mm Hg, which is greater
Recorder(optional)
Print Mode thermal array
Recording Width 48mm
Paper Speed 25mm/s, 50mm/s
Quantity of Printing Waveform dual-channel
Physical characteristic:
Requirement of power AC 100~240V, 50/60Hz
Power consumption 150VA
Dimension 310×140×263(mm)
Net weight 3.8Kg