Medical Diagnosis Equipment
Hospital Five Part Hematology Analyzer (HA-7000)
2017-10-25 18:21  Visit:170
Price:Unfilled
Send inquiry
5 part differentiation,
 29 parameters,
 3D scatter gram , 

Principles
Impedance method for WBC, RBC, PLT counting
Cyanide-free reagent for Hemoglobin test
Flow Cytometry (FCM), Semi-conductor Laser scatter, chemical dye method, independent Basophil channel
 
Parameters
WBC, Lym%, Lym#,Mon%, Mon#,Neu%, Neu#,Eos%, Eos#, Bas%, Bas#, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, PDW, MPV, PCT,and 4 research parameters P-LCC , P-LCR, LIC%, LIC#, ALY%, ALY#
1 3D and 3 2D 3 scatter grams and 3 histograms 
 
Throughput
Up to 60 samples per hour
 
Sample Volume
Whole Blood: 20ul
Prediluted: 20ul
 
Test Mode
CBC CBC+DIFF
 
Performance
Carryover  Repeatability Linearity
WBC ≤0.5% ≤2.0% (4-15×109/L) 0.00-300×109/L
RBC ≤0.5% ≤1.5% (3.5-6.0×1012/L) 0.00-8.5×1012/L
HGB ≤0.6% ≤1.5% (110-180g/L) 0-250g/L
PLT ≤1.0% ≤4.0% (150-500×109/L) 0.00-3000×109/L
 
Data Storage Capacity
Up to 100,000 results including numeric and graphical information
 
Communication
LAN port supports HL7 protocol
 
Operating Environment
Temperature: 15oC-30oC
Humidity: 30-85%
Air Pressure: 70-106 kPa
 
Power Requirement
A.C. 100-240V ≤300VA 50/60Hz
 
Dimension and Weight
380mm(L) × 450mm(W) × 535mm(H)
Weight: 45kg

Message

TO: poweam-medical

E-mail:

Content:

Contact information
Comment